Certitest™ 20 Plus Integrity Tester
Couldn't load pickup availability
Technical Parameters
| Dimensions | 495* 330* 265 mm (Length * Width * Height) |
| Weight | 13 kg |
| Operating system | Windows 11 |
| Hard disk drive | 512 G |
| Screen |
Type: Capacitance touch screen Size: 10.1 inch Screen resolution: 1280*800 |
|
Printer
|
Thermal/ Needle dot matrix |
| Power requirements | 100-240 V/AC |
| Over-voltage protection | Yes |
| Operating environment | 0-40℃;10-80% RH Single-use bags |
| Max inlet pressure | 3000-8000 mbar |
Test Methods
Bubble Point (BP)
Combined diffusion and bubble point (DF+BP)
Diffusion (DF)
Water flow test (WFT)
Pressure decay (PD)
| Measuring Ranges |
BP : 100-7000 mbar DF: 0.1-1000 mL/min WFT: 0.005-100 mL/min PD: 1-7000 mbar |
| Measuring Accuracy |
BP : ±50 mbar DF: ±4% or 2 mL/min, whichever is greater WFT: ±5% or 0.02 mL/min, whichever is greater PD: ±5% or 3.5 mbar, whichever is greater |
| Resolution |
BP : 0.01 mL/min DF: 0.001 mL/min WFT: 1 mbar PD: 1 mbar |
| Applicable Products | Membrane/ Filter/ Cassette/ Hollow fiber |
| Export Data |
Export tool: U disk Export file format: PDF |
| Data capacity |
Test data result storage: > 20 000 tests Test program storage: > 1000 programs Active user storage: > 1000 users |
| Audit trail | Compliance to 21 CFR Part 11 and GAMP standard |
| Communication ports |
1× ethernet port (RJ45) 1× USB (standard) |
Components
| Certitest 20 Plus integrity tester |
CT20PA0 (Thermal printer) CT20PA1 (Needle dot matrix printer) |
| Inlet gas tube | Ø6 x 4 mm x 2 m with quick coupling (female/open end) |
| Outlet gas tube | Ø6 x 4 mm x 2 m with quick couplings(female/female end) |
| Quick clamp | TC 50 mm to quick coupling(male end) |
| External valve | External valve with quick couplings (male/female end) |
| Case for transportation | PP trolley case |
- Cobetter_Certitest_20_Plus_Integrity_Tester_2025 Download