0.5mm Mini Hollow Fiber TFF mPES Membrane Ultrafiltration
Membrane PESMWCO 3/5/10/30/50/100/300/500/750kDaFiltration Area 28cm²/ 56cm²Package 1/pk -
1mm Mini Hollow Fiber TFF mPES Membrane, Ultrafiltration & Microfiltration
Membrane PESMWCO 3kDa~0.65μmFiltration Area 28cm²/ 56cm²Package 1/pk -
1mm Mini Hollow Fiber TFF mPES Membrane, Autoclavable
Membrane PESMWCO 750kDa-R/0.2μm-RFiltration Area 28cm²/ 56cm²Package 1/pk -
0.5mm MiniLab Hollow Fiber TFF mPES Membrane Ultrafiltration
Membrane PESMWCO 3/5/10/30/50/100/300/500/750kDaFiltration Area 118cm²/ 236cm²Package 1/pk -
1mm MiniLab Hollow Fiber TFF mPES Membrane, Ultrafiltration & Microfiltration
Membrane PESMWCO 3kDa~0.65μmFiltration Area 94cm²/ 188cm²Package 1/pk -
1mm MiniLab Hollow Fiber TFF mPES Membrane, Autoclavable
Membrane PESMWCO 750kDa/0.2μmFiltration Area 75cm²/ 150cm²Package 1/pk
Coetter hollow fiber filters are available with a broad range of sizes for easy scale-up from small volume R&D applications, small volume pilot to commercial production.
Features
· A broad range of sizes for easy scale-up
· Hydrophilic modified PES membrane provides low protein binding
· Cost-efficient and time saving
· Quick installation and easy operation, no holder required
Applications
· Purification, concentration and diafiltration of viral vector
· Clarification of cells and bacterial from fermentation broth
· Clarification of cell lysates or cell debris
· Concentration and diafiltration of proteins
· Perfusion applications
How to use a hollow fiber membrane filter?
Module Preparation
For lab and pilot applications, you will need a pump, a feed tank, a permeate collection tank, pressure gauges and valves and suitable tubing and connectors.
Cobetter hollow fiber modules made with hydrophilic modified PES membrane contain glycerin to preserve the pore structure. Glycerine is water soluble and easy to rinse out.
- Rinse out glycerine with DI water with an inlet pressure of 2-5 psi, the flushing volume should be greater than 20L/m².
- Drain module and perform module integrity test.
Note: If sanitization is necessary, use 0.1N NaOH solution as a disinfectant.
Module Integrity Test
All hollow fiber modules are integrity tested at our site. However, Cobetter highly recommends an integrity test before use. Cobetter recommends two test methods, one of which may be applied according to actual operating conditions.
- Diffusion Flow Method
- Pressure Retaining Method
Processing Buffer Equilibrium
The purpose of flushing hollow fiber modules with buffer is to ensure that the pH and salt ion conditions of the system are as similar as possible to the initial state of the sample, so as to reduce pollution and product loss.
Cleaning after Use
The hollow fiber modules must be cleaned before storage and reused by flushing with buffer or DI water, and then washed with appropriate sanitizing agent. 0.5M NaOH solution may be used to circulate for 30 min at 20-40°C.
Hollow Fiber Storage
The hollow fiber modules may be stored in 0.1N NaOH solution. For long-term storage between 1-12 months, it is recommended to be stored under refrigeration at 4°C.
Related Articles
The Applications of Hollow Fibers in Upstream Bioprocessing
mRNA Vaccine Manufacturing Strategy 1 --Plasmid DNA Production
mRNA Vaccine Manufacturing Strategy 2 - IVT in mRNA Production
The Applications of Hollow Fibers in Upstream Bioprocessing
Cobetter Minilab TFF Tangential Flow Filtration System
Application of Ultrafiltration in Recombinant Human Insulin Production