Orange Hydrophilic MCE Syringe Filters Non-sterile 100pcs/pk

In stock
Part No. SFMMCE-2213
Regular price $31.50 USD
Regular price $35.00 USD Sale price $31.50 USD (-10%)
Diameter
Pore Size
Package

Loading locations...
*Bulk Order? Request a Quote for Big Savings Now
*International Shipping within 8 Business Days
*Read our Shipping Policy and Refund Policy
*Not for medical use

The membrane material of MCE syringe filter is mixed cellulose ester which is composed of cellulose acetate and nitrocellulose, and the surface is smoother and more uniform. Suitable for filtration of water-based solvents, not resistant to strong acid, strong alkali solutions, and most organic solvents. Ideal for aqueous sample filtration that needs higher flow rate and larger volume.

Specifications

Connections Inlet Female Luer Lock
Outlet Male Luer Slip
Materials Housing
PP
Membrane  Mixed Cellulose Esters
Wettability Hydrophilic
Physical Properties Pore Size 0.22 μm, 0.45 μm, 0.8 μm
Filter Diameter
13mm 25mm 33mm
Filtration Area
0.72cm² 3.4cm² 4.5cm²
Maximum Operating Pressure
6bar 3bar 2bar
pH Range 4.0 – 8.0
Sterility Delivery Condition Nonsterile
Sterilization Method Not Recommended
Packagestrong> Pack Size 100pcs
Packaging Bulk packed

Performance

Filter Diameter Pore Size Flow Rate (mL/min@0.21Mpa,
20°C, 210kPa, 30psi)
Bubble Point
(MPa, Water)
13mm 0.22μm 64 ≥0.3
0.45μm 88 ≥0.2
0.8μm 154 ≥0.1
25mm 0.22μm 208 ≥0.3
0.45μm 409 ≥0.2
0.8μm 662 ≥0.1
33mm 0.22μm 338 ≥0.3
0.45μm 658 ≥0.2
0.8μm 846 ≥0.1


Features

  • High porosity and large water flux
  • Good interception effect
  • Excellent biocompatibility
  • High protein binding
  • Cannot be used for protein extraction

Applications

  • Sterilizing filtration of biological solutions
  • Clarification or sterilization of aqueous solutions
  • Particulate analysis and removal
  • Air monitoring and microbial detection
  • Tissue culture, biochemical analysis filtration
  • Solvent filtration which used in HPLC

Select by Volume

Sample Volume Filter Diameter Filtration Area Hold-up Volume
10mL 13mm 0.72 cm² 25μl
10mL to 100mL 25mm 3.4 cm²  < 100μl
100mL to 200mL 33mm 4.5 cm²  < 125μl

How to use a syringe filter?

Choose the right type of membrane material based on the properties of your sample. Be careful of the chemical compatibility to avoid extractables and leachables. Read this article about How to choose a syringe filter.

1. Load the sample into the syringe.

2. Attach the filter to the luer-lock of the syringe with a twisting motion. Make sure it is securely attached.

3. Filter the sample. Hold the syringe and filter vertically, and gently press the plunger to push the sample through the filter. Take care not to apply too much force or you may damage the filter.

4. Discard the used filter and repeat for the next sample.

  • Chemical_Compatibility_Chart_for_Syringe_Filters Download